MedtronicThe corporate that develops, manufactures and sells medical remedy gadgets has voluntarily recalled its MiniMed 600 Collection or 700 Collection insulin pumps as a result of the pump’s battery life could also be diminished.
The corporate started issuing area operations on July 31, notifying prospects worldwide that they need to monitor the built-in alerts and alerts on MiniMed pumps to trace battery standing and get in touch with the corporate in the event that they discover a change within the pump’s battery life.
Medtronic stated it started analyzing the battery lifetime of its pumps after receiving complaints from prospects, saying that if the pumps have been bumped, dropped, or subjected to any bodily impression even as soon as, their battery life both decreased instantly after the drop or over time. Over time, battery life is shortened attributable to inner injury to the machine.
“This electrical concern also can trigger a battery alarm when the battery life remaining is lower than acknowledged within the consumer information. This will trigger the pump to cease delivering insulin prior to usually anticipated and should end in hyperglycemia or diabetic ketoacidosis (DKA ) and different well being dangers might require medical intervention,” the corporate stated in a press release.
“Most lately, from January 2023 to September 2024, Medtronic acquired 170 studies of hyperglycemia >400 mg/dL and 11 studies of diabetic ketoacidosis in the US which may be associated to this drawback.”
The FDA categorised the problem as a Class I recall, probably the most severe sort of recall, which means there’s a excessive chance that use of the product will trigger severe damage or loss of life.
The corporate stated the pump will nonetheless generate battery alerts and alarms, however these alarms and alarms is not going to be delivered “when the battery has regular remaining life and should be changed to keep away from the pump stopping insulin supply.”
Medtronic says customers ought to change the batteries as quickly because the “low battery” alarm goes off, and in the event that they have to be changed prior to anticipated, they need to carry an additional set of batteries and have back-up remedy if wanted.
The Dublin-based firm stated customers whose battery life has been considerably diminished, no matter whether or not their pump has skilled a drop or different bodily impression, ought to contact the corporate to find out if their pump must be changed.